SE-3&SE-300 Series Electrocardiograph User Manual Settings
- 48 -
8.3 Record Info Setup
8.3.1 Setup 1
Item
Description
Record Style
Choose from: 3Ch/3Ch, 3Ch/2Ch, 1Ch+1R, 1Ch or 3Ch+1R, the
default value is 3Ch/3Ch
Select a style to print the ECG waves of 12 leads in the auto mode.
When Record Style is set to 3Ch/3Ch, ECG waves of all leads will be
printed in 4 groups of 3.
When Record Style is set to 3Ch/2Ch, ECG waves of lead I, II, III, aVR,
aVL and aVF will be printed in 2 groups of 3, and ECG waves of lead V1,
V2, V3, V4, V5 and V6 will be printed in 3 groups of 2.
When Record Style is set to 1Ch+1R, ECG waves of all leads will be
printed one by one in a sequence, with ECG waves of one rhythm lead on
the bottom of the ECG reports.
When Record Style is set to 1Ch, ECG waves of all leads will be printed
one by one in a sequence.
When Record Style is set to 3Ch+1R, ECG waves of all leads will be
printed in 4 groups of 3, with one rhythm lead on the bottom of the ECG
reports.
NOTE: Record Style is corresponding with Work Mode.
Auto Record
Sequence
Choose from: Sequential or Synchronous, the default value is
Sequential
Select Sequential, the lead group is printed one by one in a certain

Table of Contents