SE-3&SE-300 Series Electrocardiograph User Manual Sampling and Printing ECG
- 35 -
2) Press the SENS key to set the sensitivity.
3) Set the suitable speed value and filter value.
4) Or press the MENU key to open the System Setup Screen to set the detailed settings. Press
the Return to return to the Main Screen.
5) Press the PRINT/STOP key to begin printing ECG reports. It will stop automatically after
printing a complete ECG report of 12 leads.
Press PRINT/STOP during the printing course to stop printing ECG reports. Then when you
begin printing ECG reports again, the system will print ECG reports from the first lead group.
6.2.2 MANU Mode
In the MANU mode, you need to switch lead groups by pressing the Lead switch key.
1) Press the MODE/RST key to select the AUTO mode, which will be displayed in the upper
right corner of the LCD screen.
2) Press the SENS key to set the sensitivity.
3) Set the suitable speed value and filter value.
4) Or press the MENU key to open the System Setup Screen to set the detailed settings. Press
the Return to return to the Main Screen.
5) Press the Lead switch key to select the lead group to be printed.
6) Press the PRINT/STOP key to begin printing reports.
7) Press the Lead switch key to switch the lead group while printing ECG reports and the
system will automatically print 1mV calibration mark. Press the 1mV/COPY key to print
out 1mV mark in the ECG report.
8) Press the PRINT/STOP key to stop printing ECG reports.
6.2.3 RHYT mode
In the RHYT mode, you can choose single lead type or three-lead type.
1) Press the MODE/RST key to select the RHYTHM mode; which will be displayed in the
bottom right corner of the LCD screen.
2) Press the SENS key to set the sensitivity.
3) Set the suitable speed value and filter value.
4) Press the PRINT/STOP key, the hint information Sampling will be displayed on the Main
Screen, and the sampling time will be counted. When the preset sampling time reaches, it
begins to print ECG reports.