SE-3&SE-300 Series Electrocardiograph User Manual Sampling and Printing ECG
- 33 -
Chapter 6 Sampling and Printing ECG
6.1 Entering Patient Information
After switched on, the electrocardiograph enters the main screen.
1. Configure the Patient Information Setup window. (Optional)
1) Press MENU to enter System Setup Screen.
2) Choose Patient Inof tab and press MENU to enter the sub-window.
3) Configure according to actual use. See more information in “Patient Info Setup”.
4) After configuration, press MENU or PRINT/STOP to save and quit the current
window.
5) Press Return or RECALL to return to the main screen.
2. Move the cursor to the patient symbol. Press the Left or Right arrow to enter the patient
information window on the Main Screen.
1) Press the Up or Down arrow to select the ID textbox, and press the Left or Right arrow
to pop up the editing window.
2) Press the Left or Right arrow to choose letter or number, and press MENU to confirm
the operation.
3) If the content need to be modified, press RECALL to delete.
4) After editing, press MENU or PRINT/STOP to confirm the operation and return to the
main screen.
NOTE: The patient information cannot be set or changed during the printing course.
Patient Gender (Male/Female/Empty)
Age Unit: Years, Months, Weeks or Days