Chapter 7 Managing ECG Records
If you want to save the ECG records, you should set the Auto Save to To ECG or To USB. The
default value is To ECG. Then the ECG records will be saved in the File Manager or in the U
disk automatically.
Press the RECALL key to open the File Manager where patient records are displayed. The File
Manager allows records to be stored, deleted, printed and transmitted. When there is no space for
more records to be stored in the File Manager, the message MemFull will be displayed.
7.1 Transmitting ECG Records to the PC
NOTE: To transmit ECG records to the PC, Smart ECG Viewer software of the
manufacturer must be installed in the PC. You should log into the Smart ECG
Viewer software before the transmission.
7.1.1 Transmitting ECG Records through the Serial Port
Connect the RS232 socket of the PC to the RS232 socket of the electrocardiograph with a RS232
cable. If the PC has no RS232 socket, connect the USB socket of the PC to the RS232 socket of
the electrocardiograph by using the RS232-USB assembly.
Figure 7-1 RS232-USB Assembly