Free chlorine < 0.2 mg/l
8 °dH (degrees German hardness) = 1.4mmol/l
= 10 °e (degrees English hardness)
= 14 °f (degrees French hardness)
If there is the danger of frost, you may use a water Drystar mixture of up to 50%.
DS water can be used for cooling the pump, if the pH value corresponds to the range indicated
above.
7.5 Connecng purge gas
The pump must be operated with purge gas. The purge gas is distributed in the pump via
ow restrictors as follows:
▪
to the sha-seal on the low vacuum side
▪
into the pumping chamber (balance line)
▪
to the sha-seal on the high vacuum side.
The ow restrictors ensure that the gas ow to the sha-seal on the high vacuum side
becomes very low at ulmate pressure.
The purge gas on the sha-seals provides an air cushion under the sha-seals and thus
prevents premature wear.
Use of the correct sha-seal purge and gas-ballast requirements depend on your applicaon.
If in doubt about your applicaon seng, contact Edwards.
▪ Connect an inert gas supply to the gas system. An upstream supply pressure of 4 to 10
bar(g) is required. Ensure that your gas system can maintain pressure whilst owing up
to 62 SLM of purge gas.
▪ Adjust the regulator on the gas panel to 2.2 bar(g).
▪ The purge supply should be maintained whenever the pump is operaonal, to ensure
the process seals operate correctly and that the oil boxes are adequately protected
from process media.
▪ Be aware that the purge supplies are not automacally controlled, therefore it could be
possible to pressurise the pump or system if purge is le on with a closed system.
▪ Atmospheric gas-ballast can be adjusted or shut o completely, depending on your
applicaon, using the manual needle valve.
WARNING: LEAKAGE OF GAS
The atmospheric sha-seal between the gearbox and atmosphere is a PTFE dynamic seal,
which may, under wear-out condions, lead to small leakage of gearbox gas. Ensure that
purge gas is supplied connuously to the main pump sha-seals to prevent process gas
entering the gearbox which could otherwise leak from the atmospheric sha-seal.
Leakage of purge gas to atmosphere should also be considered; ensure the pump is in a
well venlated area.
Page 29
A41802880_C - Installaon