7.6.5 Check the direcon of rotaon of the motor
Blank the inlet or connect the EDS pump to the vacuum system before you check the
direcon of pump rotaon. If you do not, there is danger of objects being trapped in the
rotang pump rotors.
Direcon of rotaon is clockwise as viewed from the motor end of the pump. View the
motor fan direcon.
Let the pump run up briey (for approximately 1 second) and then immediately switch it o
again.
If the rotaon direcon is incorrect, isolate and lock-out the power supply, then change the
phases on the pump-motor and recheck direcon.
7.6.6 Connecng the heat exchanger fan
(only for air cooled versions)
Connect the heat exchanger fan as shown.
Figure 16 Heat exchanger fan connecon
1. Heat exchanger fan connecon1. Heat exchanger fan connecon
The direcon of rotaon is indicated. Check the direcon of rotaon by viewing the fan. If
the direcon of rotaon is incorrect, the live and neutral phases are incorrectly wired.
Power supply must be intrinsically safe according to EN 61010 and the fan supply must be
fused with maximum 4 A.
7.7 Leak check aer installaon
EDS pumps are not hermecal sealed and leak ght by principle. Leak-check all relevant
connecons aer having installed the pump.