Fault Pump does not aain its pumping speed
Cause Intake screen is clogged.
Remedy Clean intake screen.
Cause Motor incorrectly connected.
Remedy Connect motor correctly.
Cause Motor stator defecve.
Remedy Contact Edwards Service.
Cause Motor rotor defecve.
Remedy Contact Edwards Service.
Cause Vacuum pump system has a gas leak.
Remedy Detect leak and seal it.
Cause Rotor gap is too great.
Remedy Contact Edwards Service.
Cause Bearing defecve.
Remedy Contact Edwards Service.