Cause Fricon resistance is too high due to contaminated bearings and/or contaminated
lubricant.
Remedy
Clean pump, respecvely perform maintenance.
Cause Lubricant level is too high.
Remedy Drain lubricant down to the correct level.
Cause Lubricant level is too low.
Remedy Top up lubricant to the correct level.
Cause Wrong lubricant lled in.
Remedy Contact Edwards service.
Cause Bearing is defecve.
Remedy Contact Edwards service.
Fault
Pump is extremely loud
Cause Bearing damage.
Remedy Repair pump.
Cause Thick parcle deposits.
Remedy Clean pump, respecvely perform maintenance.
Fault Motor power consumpon is too high
Cause Like "Pump gets too hot".
Remedy See Pump gets too hot
(on page 43).
Cause Incorrect mains voltage for the motor.
Remedy Connect the motor to the correct mains voltage.
Fault
Pump is too loud
Cause Motor stator defecve.
Remedy Contact Edwards Service.
Cause Motor rotor defecve.
Remedy Contact Edwards Service.
Cause Distances between housing and rotors is too small due to contaminaon.
Remedy Clean pumping chamber.
Cause Distances between housing and rotors is too small due to distoron of the pump.
Remedy Ax and connect the pump free of tensions.
Cause Rotors make contact with the housing.
Remedy Contact Edwards Service. Shutdown pump immediately.
Cause Rotor is running untrue.
Remedy Contact Edwards Service. Shutdown pump immediately.
Page 44
A41802880_C - Diagnose a fault