EasyManuals Logo
Home>Edwards>Water Pump>nEXT 730D

Edwards nEXT 730D User Manual

Edwards nEXT 730D
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
When the BRIM process completes, the TMP runs to full speed and normal TMP process
can r
esume.
You can check the TMP inacve me (storage duraon) and reset BRIM status via:
TIC – browsing to the „Service menu“, „Brg Inacve will display the TMP inacve
me. See TIC user manual for more details
nST2 – See nST2 user manual for more details.
The purpose of recording the inacve bearing me is to maximize the life of the installed
ball bearing. If the bearings are inacve for a long me, the homogeneity of the grease
distribuon deteriorates, which leads to short-term local overheang of the grease in
the bearing when the pump is directly started up to nominal speed. This can lead to
decomposion or evaporaon of the grease. This grease is then no longer available for
the running me of the bearing and ulmately leads to earlier failure of the ball bearing.
The Bearing Run In Mode (BRIM), on the other hand, restores the homogeneity of the
grease distribuon by gently restarng the pump.
6.4 Bearing run in mode
Bearing run in mode (BRIM) is a start-up sequence for the pump designed to
precondion the bearing to opmise bearing life. The pump goes through a number of
speed steps before reaching full speed.
The BRIM process last around 4.3 hours and the LED indicates when this is running.
BRIM mode can be cancelled by sending a stop signal and then start signal to a pump
currently running the BRIM process. This is NOT recommended and may invalidate
the warranty.
BRIM mode is either triggered:
automacally by the pump
manually thr
ough nST2
manually through the TIC, or
by pressing both speed adjustment push buons (loca
ted behind the service
cover) simultaneously for 2s on the pump.
Table 1
Bearing LED
LED is not lit while pump is powered on No bearing condioning required
LED is ashing 0.25 s ON and 0.75 s OFF Pump has been stored for more than 12 months,
BRIM is required and will start on the next pump
start
LED is permanently ON BRIM is running
LED is ashing 0.25 s ON and 0.25 s OFF Pump has been stored for more than 3 years, bear-
ing replacement is required
06/2021 - ©Edwards Limited
Page 27300872820_002_C2
B8J203880_C
300872820_002_C2 - Assembly

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards nEXT 730D and is the answer not in the manual?

Edwards nEXT 730D Specifications

General IconGeneral
BrandEdwards
ModelnEXT 730D
CategoryWater Pump
LanguageEnglish

Related product manuals