EasyManuals Logo

Edwards nXR60i User Manual

Edwards nXR60i
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
4.5.2 Connect the electrical supply
WARNING: PROTECTIVE EARTH CONNECTION
Risk of electric shock. Mak
e sure that the pump and electrical cables are suitably
protected against earth (ground) faults. We recommend you aach a protecve earth
(ground) conductor (with a cross seconal area of 2.5 mm
2
/14 AWG) to the
protecve earth (ground) stud.
Make the electrical c
onnecon to the pump mains connector port (Figure: nXRi pumps,
(10)) with an appropriate cord set. Refer to Table: Recommended cord sets.
The pump must be grounded through the conductor of the mains input connector.
4.6 Connecon for remote control and monitoring
To operate the pump with parallel or serial control, the 15-way D-type connector must
be used (Figure: nXRi pumps, (8)).
Refer to Figure: Logic interface connecons - analogue speed control for details of the
logic interface pins.
The pump is also equipped with a front USB port for service purposes using serial
communicaon. This connector cannot be used for pump operaon, but for seng and
reading parameters and for pump monitoring.
05/2021 - ©Edwards Limited
Page 24A90301880_C
A90301880_C - Ins
tallaon

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards nXR60i and is the answer not in the manual?

Edwards nXR60i Specifications

General IconGeneral
BrandEdwards
ModelnXR60i
CategoryWater Pump
LanguageEnglish

Related product manuals