CAUTION: PUMP LIFE
Risk of damag
e to equipment. Unpack the pump carefully and avoid excessive shock
to the pump. Excessive shock will damage the bearings and reduce the life of the
pump.
Refer to Table: Mechanical data for the mass of the pump.
The pump is supplied in a recyclable packaging.
1. Open the cardboard box from the top.
2. Remove the packing material. Keep all the packing materials for use in inspecon
and if the pump is returned for service.
3. Examine the pump for damage. If the pump is damaged, nofy your supplier and
the carrier immediately. Give the supplier and the carrier the informaon that
follows:
▪ part number of the pump
▪ serial number of the pump
▪ order number
▪ supplier's invoice number
4. Do not use the pump if the pump is damaged.
5. If the pump is not to be used immediately, store the pump in the condions given
in Storage on page 45.
6. The pump is supplied with sealed inlets and outlet to prevent the entry of dust and
v
apor. Do not remove the seals unl the pump is ready to be installed on the
vacuum system.
4.3 Mechanical installaon
WARNING: HEAVY OBJECT
Risk of ph
ysical injury. Use suitable equipment to li the pump.
WARNING: LOOSE CABLES/PIPEWORK
Tripping hazard. Make sure that any cables and/or pipe work aached to the pump
are xed carefully to avoid a slip/trip hazard and to prevent any damage to the cable.
WARNING: LIFTING EYE INTEGRITY
Risk of injur
y or damage to equipment. Make sure that the maximum angle between
the paired slings used to li the system is 45 degrees. Increased angle could
compromise the liing eye integrity.
05/2021 - ©Edwards Limited
Page 20A90301880_C
A90301880_C - Ins
tallaon