Note:
The s
top command does not isolate the pump from the electrical supply.
Standby
In standby mode the pump operates at a reduced speed to improve the service life of
the pump.
1. Push the standby buon to select standby mode. The pump will inially run at
factory default standby speed (70% of full speed).
2. Adjust the speed with the increase and decrease standby speed buons. The
maximum standby speed is 100% of the default run speed and the minimum
standby speed is 50% of the default run speed.
A single short push will change the speed by 1% of the default run speed. Hold the
buon to change the speed by 1% per second.
3. Once adjusted, the pump will return to the new user dened speed each me
standby speed is selected.
4. Push the standby buon to return to normal run speed.
So start and so stop modes
For most pumps operang under normal laboratory condions or at ulmate pressure
the so-stop mode or so-start mode is unlikely to be invoked.
So stop:
If the pump is operang within a high ambient temperature the product may enter a
so-stop mode when the o command is sent or stop buon is pressed, this protects
the product mechanism during the shutdown procedure by enabling the pump to relax
back into its stopped state. The so-stop mode is likely to be acvated in ambient
temperatures higher than 30 °C. During the so-stop mode the pump speed is halved
and the cooling fan speed is increased to enable controlled cooling of the mechanism.
Once a pre-dened temperature level has been met the pump will switch o completely
and is then ready for restart.
The so-stop cool down me period is expected to be < 1 minute in the majority of
cases, but could last up to 4minutes depending upon the ambient temperature and load
applied to the pump.
Once the so-stop process has been acvated the pump cannot be restarted unl the
shutdown is complete.
So start:
If the internal pump temperature is above a pre-dened threshold, the pump can not
start immediately, when the start command is sent or start buon is pressed. Instead,
the cooling fan will switch on at full speed to cool the pump and once temperatures are
suciently low the pump will run up to full speed as normal. The delay to start is
expected to be < 1 minute in the majority of cases but may last longer depending upon
the ambient temperature of the cooling air and the inial temperature of the pump
module.