18
MODEL
SE 7430
SE 7440
SE 7441
SE 7451
SE 7452
Dwell Time, Second
Auto Range
0, 0.1-999.9
(0=continuous)
0.1
±(0.1% of Setting + 0.05s)
Fixed Range
0, 0.2-999.9
(0=continuous)
Current Offset
0.000-40.00mA(Total current + current Offset 40mA) for SE 7430 & SE 7440 & SE 7441
0.000-100.0mA(Total current + current Offset 100mA) for SE 7451 & SE 7452
Arc Detection
The range is from 1-9(9 is the most Sensitive)
DC WITHSTAND VOLTAGE
Output Rating
6KVDC/10mA
6KVDC/10mA
6KVDC/10mA
6KVDC/10mA
6KVDC/10mA
Output
Rating(OPT.7015)
7.5KVDC/10mA
7.5KVDC/10mA
7.5KVDC/10mA
7.5KVDC/10mA
7.5KVDC/10mA
Output Voltage, kVdc
0-6.00
0.01
±(1% of Setting + 0.5% of Range)
Output Voltage, kVdc
(OPT.7015)
0-7.50
0.01
±(1% of Setting + 0.5% of Range)
Output Ripple
< 4%(6KV/10mA at Resistive Load)
HI and LO-Limit,
current, uA
Auto Range
0.0000-0.9999
0.0001
±(2% of Setting + 10 counts)
Low Range is ON.
1.000-9.999
0.001
10.00-99.99
0.01
100.0-999.9
0.1
±(2% of Setting + 2 counts)
1000-10000
1
Fixed Range
0.0000-0.3500
0.0001
±(2% of Setting + 10 counts)
Low Range is ON.
0.351-3.500
0.001
3.51-35.00
0.01
35.1-350.0
0.1
±(2% of Setting + 2 counts)
351-3500
1
3501 10000
1
Ramp Up Timer,
Second
0.4-999.9 (Low Range = OFF)
0.1-999.9 ( when fixed range is ON)
0.1
±(0.1% of Setting + 0.05s)
0.5-999.9 (Low Range = ON)
Ramp Down Timer,
Second
0.0, 1.0-999.9
Dwell Time, Second
0, 0.4-999.9 , (Low Range = OFF)
(0=continuous)
0, 1.0-999.9 , (Low Range = ON)
(0=continuous)
0, 0.1-999.9s, when it be select fixed
range (if low range is ON , the time will
follow low range setting)
Ramp-HI, current , uA
Low Range is ON
0.0000-0.9999
0.0001
±(2% of Setting + 10 counts)
1.000-9.999
0.001
10.00-99.99
0.01
100.0-999.9
0.1
±(2% of Setting + 2 counts)
1000-10000
1
Low Range is OFF
0.0-999.9
0.1
1000-10000
1
Charge LO, current,
uA
0.0-350.0 µA DC or Auto Set, When current range is 3 ranges.
0-3500 nA DC or Auto Set, When current range is 6 ranges.
Discharge Time
< 50 msec for no load, < 100 msec for capcitor load(all capacitance values in MAX load spec below)
Maximum Capacitive
Load DC Mode
F < 1KV, 0.75µF < 2KV, 0.5µF < 3KV, 0.08µF < 4KV, 0.04µF < 5KV, 0.015uF < 6KV