EasyManua.ls Logo

EEC SE 7441 - 4. Programming Instructions; 4.1 Using the Touch Screen

Default Icon
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20
Measurement
Range
Resolution
Accuracy
Voltage, KV(AC/DC)
0.00-7.50
0.01
±(1% of reading + 0.5% of Range)
Voltage, Vdc
(IR only)
0-1000
1
±(1% of reading + 0.5% of Range)
1001-6000
1
±(1% of reading + 0.5% of Range)
AC Total Current, mA
(Auto range)
0.000-3.500
0.001
±(2% of reading + 3 counts)
3.00-40.00
(SE 7430 , SE 7440 , SE 7441)
0.01
3.00-99.99
(SE 7451 , SE 7452)
0.01
±(2% of reading + 6 counts)
100.0
(SE 7451 , SE 7452)
0.1
AC Total Current, mA
(Fixed range)
0.000-3.500
0.001
±(2% of reading + 2 counts)
0.00-40.00
(SE 7430 , SE 7440 , SE 7441)
0.01
±(2% of reading + 2 counts)
Additional error when reading < 6%
of range for ±(0.1% of range)
0.00-99.99
(SE 7451 , SE 7452)
0.01
±(2% of reading + 2 counts)
Additional error when reading < 3%
of range for ±(0.1% of range)
AC Real Current , mA
0.000-9.999
0.001
±(3% of reading + 50µA)
All Ranges PF > 0.1 , V > 250VAC
10.00-99.99
0.01
DC Current, nA
(Auto/Fixed Range)
0.0-350.0
0.1
±(2% of reading + 10 counts)
Low Range is ON.
DC Current, µA
(Auto Range)
0.300-3.500
0.001
3.00-35.00
0.01
30.0-350.0
0.1
±(2% of reading + 2 counts)
DC Current, µA
(Fixed Range)
0.000-3.500
0.001
±(2% of reading + 10 counts)
Low Range is ON.
Additional error when reading < 6%
of range for ±(0.1% of range)
0.00-35.00
0.01
00.0-350.0
0.1
±(2% of reading + 2 counts)
DC Current, mA
(Auto Range)
0.300-3.500
0.001
3.00-10.00
0.01
DC Current, mA
(Fixed Range)
0.000-3.500
0.001
±(2% of reading + 2 counts)
Additional error when reading < 6%
of range for ±(0.1% of range)
0.00-10.00
0.01
±(2% of reading + 2 counts)
Additional error when reading < 6%
of range for ±(0.2% of range)
AC Current, A (GB)
0.00-40.00
0.01
±(2% of reading + 0.5% of Range)
Resistance, mΩ (GB)
0-600
1
3.00-32.00 A,
±(2 % of reading + 2 counts)
Resistance, Ω
(Continuity)
0.000-1.000
0.001
±(1% of reading + 3 counts)
1.01-10.00
0.01
10.1-100.0
0.1
101-1000
1
1001-10000
1
±(1% of reading + 10 counts)