36
Errors/warnings Description and potential causes Solution
Functionality test not successful
Test solution has been used more than
three months after the bottle was
opened or is past its expiry date.
Test solutions not stored correctly.
Sensors were not stored correctly or
have exceeded their expiry date.
Repeat the functionality test with new
materials after checking the code. Follow
the instructions for performing the
functionality test.
Make sure that the target concentration
you have selected in the menu matches
the concentration of the test solution.
If the problem persists, please contact
theLactate Scout 4 service team.
Electronic error
inside the device.
Turn the device o and on again.
If the problem persists, please contact
theLactate Scout 4 service team.
Storage error Please contact the Lactate Scout
4service team.