DK50-10
01/2022 - 33 - NP-DK50-10-16_01-2022-MD
Maintenance intervals
Performed
by
user
qualified professional
a)
data is presented in hours; if not possible, then the data is considered in years
b)
only for compressors without dryers
c)
For 60 Hz compressor models, reduce the time interval in hours by
20%
(2000h./1600h., 4000h./3200h., 6000h./4800h.,
8000h./6400h., 10000h./8000h., 12000h./9600h.)
d)
only for compressors with dryers
Set of spare
parts
-
-
-
-
-
-
-
02520
w0304
-
000
-
025200126
-
000
Chapter
19.2
19.5
19.5
9
19.3
19.4
19.9
19.8
19.6
19.7
19.
12000 h
9600 h
x
10000 h
8000 h
8000 h
6400 h
x
6000 h
4800 h
4000 h
3200 h
x
2000 h
1600 h
once
ever
y 2
years
x
x
once
a year
x
x
x
x
x
once
a week
x
x
once
a day
x
x
50 Hz
60 Hz
Product operational checks
Drain the condensate from the air tank
- At
high RH )
b)
- Pour the condensate out of the vessel
d)
Drai
n the
condensate from the air tank
- At normal RH )
b)
Product function check
Leak check connections and inspect
device
Inspection of electrical connections
Cooler and fan check
Filter element replacement in the dryer
filter
Safety valve check
Replacing the inl
et filter on the air pump
a)
Conduct a “repeated test” per EN 62353
Time interval
c)

Table of Contents

Other manuals for EKOM DK50-10