27
10 ERROR CODES
Press to clear error codes.
Error code Explanation
0
Err
Unable to obtain a stable reading. This is either due to a very high electrical noise
environment, beyond emc specification limits, or could indicate a gauge fault.
1
Err
Calibration factors are out of range. Clear the error message and repeat the adjustment
procedure.
2
Err
Averaging error.
3
Err
No readings taken for average calculation.
4
Err
Statistics memory full.
5
Err
Statistics overflow, more than 9,999 readings have been taken. Statistics data must be
cleared (see Section 6.3) before further readings can be taken for statistical analysis.
6
Err
Gauge over-range reading.
7
Err
Memory hardware fault. Contact Elcometer or your local Elcometer supplier.
8
Err
Gauge is not configured after repair. Contact Elcometer or your local Elcometer supplier.
9
Err
The printer is off line, out of paper or printer or cable fault. Correct printer fault or press
to cancel print function.