PR2000S SERIAL PRINTER BAUD RATE SET-UP.
Before commencing the this setting please ensure the printer is off and
disconnected from the electrical supply.
The PR2000S serial printer must be set up to make it compatible with the setting made
to the IA unit. The Baud rate on the PR2000S printer is factory set at 9600. This must
be changed to 4800. The procedure for changing the baud is as follows:
Press the „ and the „1‟ button simultaneously. The printer will now enable the
„PROGRAMME MODE‟.
The following text will be printed. “DATA/PAR‟Y:- 8/NONE”
Press the „PAPER‟ feed button.
The following text will be printed. “Xon/Xoff:- ON”
Press the „PAPER‟ feed button.
The following text will be printed. BAUD RATE:- 9600”
Press the button
The following text will be printed. “BAUD RATE:- 4800”
Press the and „PAPER‟ feed buttons simultaneously.
The PR2000S will produce a print of the stored settings which will be as follows:-
SETUP COMPLETE
:-
Data/Par‟y
:-
8/NONE
Xon/Xoff
:-
On
Baud rate
:-
4800
Country
:-
UK
Print mode
:-
Text
Auto-off
:-
5 min
Emulation
:-
Standard
DTR
:-
Normal
Interface
:-
Serial
Mechanism
:-
M190