SECTION 3: SWEAT INDUCTION AND COLLECTION
3.2 Inducing Sweat
Iontophoresis Cancelled Manually or by Error
If the iontophoresis is manually cancelled or an error condition occurs, an
audible tone sounds and Iontophoresis Cancelled is displayed and a pop
up display will appear.
1. Tap to exit from the pop-up.
2. Tap to advance to the next screen.
NOTE: If iontophoresis is cancelled, the iontophoresis procedure must
be repeated to completion before proceeding to sweat collection.
3. Remove the electrodes and discard the Pilogel discs.
Tap to advance to the next screen.
4. Clean the skin with deionized water and dry the skin.
5. Tap to advance to the next screen . or tap to go back to the
previous screen.
A Summary screen displays the operator-entered information up
through the time iontophoresis was manually stopped or the error
condition occurred. Information not entered or completed displays in
gray.
6. Tap to return to the Home screen.