What to do if ELITechGroup MACRODUCT ADVANCED 3710 SYS displays Error 1014 Ramp Timeout Low mA?
- AApril LeSep 8, 2025
If your ELITechGroup Personal Care Products displays 'Error 1014 Ramp Timeout Low mA', inspect the electrodes and clean if necessary. Place a drop of deionized water between the electrode and Pilogel disc and directly on the clean skin beneath the Pilogel disc. Make sure the electrodes are secured with adequate tightness to the patient’s limb. Suggest repeating the test once. If the problem persists, with electrodes not attached to a patient, try checking the electrodes from the System screen. (From the Home screen, tap, then tap System and then tap Functional Test.)