61
SECTION 3: SWEAT INDUCTION AND COLLECTION
3.3 Collecting Sweat
7. Insert Needle Using Syringe or Sweat Dispenser
Insert EasyDuct Needle Using the Syringe
a. Position the syringe plunger at mid-point before inserting the
EasyDuct needle into the tubing.
b. DO NOT squeeze the syringe body or move the syringe plunger at
any time while inserting the EasyDuct needle into the tubing or
during the following procedure.
c. Holding the open end of the tubing in one hand, carefully insert
the EasyDuct needle approximately 5 mm (¼ inch) into the
microbore tubing using a twisting motion.
d. Tap to advance to the next screen, or tap to go back to the
previous screen.
Insert Needle Using the Sweat Dispenser
a. DO NOT squeeze the dispenser at any time while inserting the
needle into the tubing or during the following procedure. Avoid
squeezing the dispenser while manipulating the tool during
needle insertion. STRONGLY RECOMMENDED: grasp the dispenser
over the black forward end rather than on the flexible middle
section.
b. Holding the open end of the tubing in one hand, carefully insert
the needle approximately 5 mm (¼ inch) into the microbore
tubing using a twisting motion.
c. Tap to advance to the next screen, or tap to go back to the
previous screen.
CAUTION:
Do not use the Sweat Dispenser to introduce sweat samples into the
Sweat-Chek Analyzer.