EMC INDUSTRIAL GROUP LTD Initial Setup
MW61A_IM_ALL_SV6.09d_en 23/47
Step Quick Keys
p32 Set the digital output terminal functions 243x
INFORMATION, RESETS & FINAL CALIBRATION
p35 Set key locks or customise the display 256xx
Initial Setup
Before calibration can be done, the system integrators will need to have completed all the wiring and
you will need suitable test weights.
Engineering Units
By default the measuring units of the controller are;
• weight kg
All settings must be entered using these measured units. The units can be changed if required.
Step Quick Keys
p24 Select weight units 2112
Measuring Range
Set the measuring range to the largest value that will occur during normal operation. Most settings hav-
ing the same units as one of the capacities are stored as a percentage of this capacity. If the capacity is
changed, these settings will also change.
Step Quick Keys
p24 Enter weight capacity 2121
Display Resolution
The display resolution for each engineering value can be changed if needed.
Step Quick Keys
p24 Enter weight division 2131
Calibration
Calibrate Loadcell
Step Quick Keys
p26 Calibrate loadcell 22321 remove all weights and set to 0
22322 apply test weights & enter test weight value