100-412-196 REV. 06 57
6.5.2 Time Mode
In this mode, ultrasonics are applied to the sample for a specified period of time.
Ultrasonics will remain on until the timeout period has been reached unless a STOP
condition occurs.
The following table shows the parameters of On Time mode, along the default, max. and
min. values.
Entering an illegal value will generate 3 beeps. The system will not accept out-of-range
parameters. (See
7.4 Alarms/Errors for further details).
Table 6.6 Time Mode Parameters
Parameter Default Max. Value Min. Value
On Time 0.05 s 9.99 s 0.05 s
Amplitude 50% 100% 10%
Off Time 0.05 s 9.99 s 0.05 s
Ground Detect 0.05 s 0.99 s 0.00 s
NOTICE
You can return to the ready screen without saving any change by pressing the ESC key
at any time.
Table 6.7 Time Mode Operational Sequence
Step Action Reference
1
Turn on power, wait for the LCD to show ready and current
mode.
2
Press the Mode/Configuration key once to get into mode
selection. Current selected mode icon will blink.