6.5.3 Continuous Mode
In this mode, ultrasonics are applied to the sample until you stop the cycle. The user
controls the timing of ultrasonics manually. The power supply starts when a START
condition is received and stops when a STOP condition is received.
The following table shows the parameters of Continuous mode, along the default, max.
and min. values.
Entering an illegal value will generate 3 beeps. The system will not accept out-of-range
parameters. (See
7.4 Alarms/Errors for further details).
Table 6.8 Continuous Mode Parameters
Parameter Default Max. Value Min. Value
Amplitude 50% 100% 10%
Ground Detect 0.05 s 0.99 s 0.00 s
NOTICE
You can return to the ready screen without saving any change by pressing the ESC key
at any time.
Table 6.9 Continuous Mode Operational Sequence
Step Action Reference
1
Turn on power, wait for the LCD to show ready screen and
current mode.
2
Press the Mode/Configuration key once to get into mode
selection. Current selected mode icon will blink.