94 Micro Motion Series 1000 and Series 2000 Transmitters
Pressure Compensation, Temperature Compensation, and Polling
Figure 9-5 Configuring polling with ProLink II
Polling for pressure
Set External temperature
to 32 °F (0 °C)
Apply
Polled variables
Specify Polling control
(1)
Enter HART tag
of external device
Specify Process variable
Polled variable 1
Specify Polling control
(2)
Enter HART tag
of external device
Specify Process variable
Polled variable 2
Apply Apply
DoneDone
Apply Apply
Polling for temperature
ProLink >
Configuration >
ProLink >
Configuration >
Temperature
(1) Choose Primary if the external
device will probably be accessed
by another device acting as a
secondary master (e.g., a
Communicator). Choose
Secondary if the external device
will probably be accessed by
another device acting as a
primary master.
(2) If you are configuring both Polled
Variable 1 and Polled Variable 2,
use the same Polling Control
setting for both. If you do not, Poll
as Primary will be used for both
devices.