EasyManua.ls Logo

Emerson Micro Motion 1000 Series

Emerson Micro Motion 1000 Series
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuration and Use Manual 95
Pressure Compensation, Temperature Compensation, and Polling
Measurement Performance TroubleshootingCustody TransferCompensation
Figure 9-6 Configuring polling with the Communicator
9 Ext temp
Set Static temperature
to 32 °F (0 °C)
Send
8 Polling setup
Enter Ext Dev Tag 1
Specify Polled Var 1
Specify Poll Control 1
(1)
Done
Send
1 Charize Sensor
Home
Enter Ext Dev Tag 2
Specify Polled Var 2
Specify Poll Control 2
(2)
Done
Send
Home
1 Charize Sensor
Polling for pressure Polling for temperature
On-Line Menu >
5 Detailed Setup
On-Line Menu >
5 Detailed Setup
(1) Choose Primary if the external
device will probably be accessed by
another device acting as a secondary
master (e.g., a Communicator).
Choose Secondary if the external
device will probably be accessed by
another device acting as a primary
master.
(2) If you are configuring both Polled
Variable 1 and Polled Variable 2, use
the same Poll Control setting for
both. If you do not, Poll as Primary
will be used for both devices.

Table of Contents

Related product manuals