164 Micro Motion Series 1000 and Series 2000 Transmitters
Default Values and Ranges
Meter factors Mass factor 1.00000
Density factor 1.00000
Volume factor 1.00000
Density Density damping 1.6 sec 0.0–51.2 sec User-entered value is
corrected to nearest value in
list of preset values.
Density units g/cm
3
Density cutoff 0.2 g/cm
3
0.0–0.5 g/cm
3
D1 0.00000
D2 1.00000
K1 1000.00
K2 50,000.00
FD 0.00000
Temp Coefficient 4.44
Slug flow Slug flow low limit 0.0 g/cm
3
0.0–10.0 g/cm
3
Slug flow high limit 5.0 g/cm
3
0.0–10.0 g/cm
3
Slug duration 0.0 sec 0.0–60.0 sec
Temperature Temperature damping 4.8 sec 0.0–38.4 sec User-entered value is
corrected to nearest lower
value in list of preset values.
Temperature units Deg C
Temperature calibration factor 1.00000T0.0000
Pressure Pressure units PSI
Flow factor 0.00000
Density factor 0.00000
Cal pressure 0.00000
T-Series sensor D3 0.00000
D4 0.00000
K3 0.00000
K4 0.00000
FTG 0.00000
FFQ 0.00000
DTG 0.00000
DFQ1 0.00000
DFQ2 0.00000
Special units Base mass unit g
Base mass time sec
Mass flow conversion factor 1.00000
Base volume unit L
Base volume time sec
Volume flow conversion factor 1.00000
Table A-1 Transmitter default values and ranges continued
Type Setting Default Range Comments