Terminate the 4-wire cable drain wires only at the sensor/core processor end of the
cable. See the sensor installation manual for more detail. Do not connect the 4-wire
cable drain wires to the ground screw located inside the 5700 junction box.
See Figure 3-10 for 4-wire terminal connections.
See Figure 3-11 for 9-wire terminal connections.
Figure 3-10: 4-wire transmitter-to-sensor wiring connections
Figure 3-11: 9-wire transmitter-to-sensor wiring connections
Note
Connect the 4 drain wires in the 9-wire cable to the ground screw located inside the
junction box.
4. Replace the transmitter-to-sensor wiring compartment cover and tighten the
screws to 14 in lbf (1.58 N m) to 15 in lbf (1.69 N m).
Installation Manual Mounting and sensor wiring
MMI-20027478 February 2020