1 Startup procedure
Complete the following steps to start the 5081 transmitter.
1. Using the infrared remote control (IRC), press PROG, NEXT, NEXT, and ENTER in this
order.
2. Select the measurement type and unit of measurement if using a 5081-P or 5081-C.
3. Use the arrow keys to toggle between Celsius and Fahrenheit.
4. Press ENTER and then RESET.
5. Press PROG, NEXT, and ENTER in this order.
6. Use the arrow keys to toggle T AUTO between ON or OFF. This determines whether
the transmitter uses the process temperature (ON) or a manual temperature (OFF).
Once the selection has been made, press ENTER. If OFF is selected, then enter the
manual temperature desired using the arrow keys. Once completed, press ENTER.
Startup procedure
Instruction Manual 1