2. Press ENTER.
T REF appears.
3. If needed, change the reference temperature from the factory default 77 °F (25 °C)
to a different reference temperature for the process. Press ENTER.
UNIT appears.
4. Press the up or down arrow to select the desired measurement units: uS
(microsiemens), mS (millisiemens), none (no units displayed), % (percent), or ppm
(parts per million); then press ENTER.
NUM PTS appears.
5. Press up or down arrows to select the desired number of data points for a custom
conductivity curve. Selecting 2 will generate a linear relationship for conductivity
and concentration at the given reference temperature.
a. Enter the concentration for Pt. 1 (displayed as uS 1). Press ENTER.
b. Enter the known conductivity for Pt. 1 in uS/cm. Press ENTER.
c. Complete this process for additional known data points. Press ENTER.
CALC CUST appears briefly.
d. Press ENTER.
PROCESSING appears briefly; then APPLY CUST appears.
e. Press ENTER to register the custom curve into memory and return to the SETUP
CUST screen.
The custom curve will now be used to display and output all conductivity measurement
when CUST is selected in the display menu for measurement type.
6.5 Restore to default factory settings
6.5.1 Configuration overview
All user-defined configurations can be erased to return the transmitter to the factory
default settings.
6.5.2 Returning to factory default settings
1. Press PROG on the remote controller.
2. Press NEXT until DEFAULT appears in the display. Press ENTER.
3. Use the up and down arrows to toggle between NO and YES. Press ENTER to return
to the factory default settings.
Programming - basics
Instruction Manual 37