EasyManua.ls Logo

Emerson Rosemount5081 - Systematic Troubleshooting; RTD Resistance Values

Emerson Rosemount5081
152 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3. To troubleshoot the transmitter independently of the sensor, use an appropriate
resistor across the temperature input connectors and connect the conductivity
inputs to resistance decade box.
8.2.5 Systematic troubleshooting
If the quick troubleshooting guide does not resolve the error, try the step-by-step
approach offered in in the flowchart below.
1. Follow the troubleshooting flowchart.
2. Refer to the tests and instructions indicated by the flowchart to diagnose the
problem.
8.2.6 RTD resistance values
Table 8-5 is a ready reference of RTD resistance values at various temperatures. These are
used for testing and evaluation of the sensor.
NOTICE
Resistance values are read across the RTD element and are based on the manufacturer's stated
values (± 1%). Allow enough time for the RTD element in the sensor to stabilize to the
surrounding temperature.
RTD resistance valuesTable 8-5:
Temperature Pt-100 resistance (ohms) Pt-1000 resistance (ohms)
0 °C (32 °F) 100.00 1000
10 °C (50 °F) 103.9 1039
20 °C (68 °F) 107.79 1078
25 °C (77 °F) 109.62 1096
30 °C (86 °F) 111.67 1117
40 °C (104 °F) 115.54 1155
50 °C (122 °F) 119.40 1194
60 °C (140 °F) 123.24 1232
70 °C (158 °F) 127.07 1271
80 °C (176 °F) 130.89 1309
90 °C (194 °F) 134.70 1347
100 °C (212 °F) 138.50 1385
Diagnostics and troubleshooting
86 Rosemount 5081

Table of Contents