EasyManua.ls Logo

Emerson Rosemount5081 - SLOPE Err lo

Emerson Rosemount5081
152 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
If the message remains, go to Step 4.
4. Confirm that the problem is with the remote preamplifier. Wire the pH sensor
directly to the transmitter. Change the menu from PAMP=SnSr to trAnS for the test
and return it to SnSr afterwards.
If the error message clears, the remote preamplifier is faulty. Replace the
preamplifier.
If the error message remians, the sensor has failed. Replace the sensor.
SLOPE Err LO
SLOPE Err LO means that a two-point buffer calibration attempt has failed. The slope is too
low (<40 mV/pH) for a good measurement.
Troubleshooting SLOPE Err LO
1. Repeat the calibration.
a. Inaccurate buffers can cause a low slope. Repeat the calibration using fresh
buffers. Alkaline buffers, pH 10 or greater, are particularly susceptible to
changing value in air or with age. If a high pH buffer was used in the failed
calibration, try a lower pH buffer when repeating the calibration. For example,
use pH 4 and 7 buffer instead of pH 7 and 10 buffer.
b. Allow adequate time for readings in buffer to become constant. If the sensor was
in a process substantially colder or hotter than the buffer, allow at least 20
minutes for readings in the buffer to stabilize. Alternatively, place the sensor in a
container of water at ambient temperatures for 20 minutes before starting the
calibration.
c. Be sure the correct buffer values are being entered during calibration.
If the second calibration was successful, an error was made during the first
attempt.
If the second calibration fails, go to Step 2.
2. Check wiring. Connections to TB-10, TB-7, and TB-8 are particularly important.
Recalibrate the sensor.
if the wiring was the only problem, the sensor should calibrate.
If the message persists, go to Step 3.
3. Inspect and clean the senor. Recalibrate the sensor.
If the sensor was dirty, it should calibrate after cleaning.
If the message persists, go to Step 4.
4. Check for a faulty sensor.
If a spare sensor is available, connect it to the transmitter. Use the auto
calibration procedure to calibrate the sensor.
- If the new sensor cannot be calibrated, the transmitter is faulty. Go to Step 5.
- If the new sensor can be calibrated, the old sensor has failed.
If a spare sensor is not available, measure the glass impendance (GIMP). See
Section 8.1.7.
Diagnostics and troubleshooting
Instruction Manual 65

Table of Contents

Related product manuals