Fault removal multi X 2500
108
50 Firmware reset
Cause Remedy
Internal computer (firmware) restarted Acknowledge message
Initialize
62
64
65
66
67
Incomplete command from PC
Command from PC without STX
Command from PC: CRC error
Command from PC is invalid command
Command from PC: Invalid MEASURING command
Command MTXT from PC is missing
Cause Remedy
Communication error Acknowledge message
Initialize
130 Cl sensor: no connection
Cause Remedy
Communication faulty after detector was detected
during initialization
Acknowledge message
Initialize
131 Cl sensor: wrong command structure
Cause Remedy
Error in communication with chlorine module Acknowledge message
Switch chlorine module on/off
Initialize
132 Cl sensor: Indication error
Cause Remedy
Indicator value after titration start out of range
(measurement cannot start)
Acknowledge message
Initialize
Execution of the end point routine
Check status of measuring cell via S
YSTEM
COMPONENT TEST
CHLORINE
133 Cl sensor: wrong cell type
Cause Remedy
No initialization after cell change Initialize
134 Cl sensor: wrong status
Cause Remedy
Communication error Acknowledge message
Initialize
Check titration cell status via
menu item S
YSTEM COMPONENT TEST tab CHLORINE
Initialize
Check status of titration cell via
menu item