AOX determination multi X 2500
56
6. Insert the combustion tube into the combustion furnace.
When doing this, make sure that the connections on the
combustion tube fit into the recesses on the furnace.
7.
Attach the oxygen sluice (1) to the top of the combustion
tube.
Connect the angled adapter (2) with the oxygen hose (hose
no. 3) to the oxygen sluice.
NOTICE! Do not push the oxygen hose and the connector
beyond the branch length of the angled adapter on the sluice
(see arrow). Only then, the gas will be able to flow freely.
The open combustion tube is mounted and can be
connected to the sulfuric acid container.
8. Mount the heat bridge in the cutout for the furnace above
the gas outlet of the combustion tube.
Insert the sulfuric acid container (vertically):
WARNING
Concentrated sulfuric acid can cause severe chemical burns! Wear appropriate
protective clothing when working at the sulfuric acid container. Obser
ve all
specifications in the safety data sheet!
In the standard configuration, the sulfuric acid container is attached to the furnace of
the multi X 2500 and directly connected to the open combustion tube via a spherical
joint.