multi X 2500 AOX determination
57
1. Attach the drip pan (1) to the combustion furnace.
2.
Fill the sulfuric acid container (vertical, 2) with 40 ml of
concentrated sulfuric acid.
3.
Insert the plug to the sulfuric acid container.
4.
Carefully place the sulfuric acid container in the drip pan of
the combustion furnace.
5.
Connect the sulfuric acid container to the open combustion
tube using the forked clamp (3).
6. Connect the measuring gas hose (4) via the spherical joint
adapter to the sulfuric acid container. Secure the connection
with a forked clamp.
7.
If not already done, route the measuring gas hose through
the holes at the top of the basic device's and the chlorine
module's side panels.
8. Put the top cover into place and close the doors.
The sulfuric acid container (vertical) is mounted.