Technical data Cerabar PMC11, PMC21, PMP11, PMP21, PMP23
40 Endress+Hauser
12.4.6 Thermal change of the zero output and the output span
PMP11, PMP21
Measuring cell –20 to +85 °C (–4 to +185 °F) –20 to –40 °C (–4 to –40 °F)
+85 to +100 °C (+185 to +212 °F)
% of the calibrated span for TD 1:1
<1 bar (15 psi) <1 <1.2
≥ 1 bar (15 psi) <0.8 <1
PMP23
Measuring cell –10 to +85 °C (+14 to +185 °F) +85 to +100 °C (+185 to +212 °F)
% of the calibrated span for TD 1:1
<1 bar (15 psi) <1 <1.2
≥ 1 bar (15 psi) <0.8 <1
12.4.7 Long-term stability
1 year 5 years 8 years
% of URL
±0.2 ±0.4 ±0.45
12.4.8 Switch-on time
≤2 s