Operation Deltabar S FOUNDATION Fieldbus
30 Endress+Hauser
6.3 FOUNDATION Fieldbus interface
See operating instructions.
6.3.1 Device identification and addressing
FOUNDATION Fieldbus identifies the device using its ID code and automatically assigns it a
suitable field address. The identity code cannot be changed.
The device appears in the network display once you have started the FF configuration program
and integrated the device into the network. The blocks available are displayed under the device
name.
If the device description has not yet been loaded, the blocks report "Unknown" or "(UNK)".
Deltabar S reports as follows:
P01-xxxxxxxx-19-xx-xx-xx-134
– DIP switch 1: for locking/unlocking parameters relevant to the measured value.
Factory setting: off (unlocked)
– DIP switch 2: for the simulation mode Factory setting: off (simulation mode off)
Key(s) Meaning