6.4.4 Accepting pressure present at device as value
Example: performing position adjustment.
6.5 HistoROM
®
/M-DAT (optional)
See operating instructions.
6.6 FieldCare
See operating instructions.
6.7 Locking/unlocking operation
See operating instructions.
Onsite display Operation
P01-xxxxxxxx-19-xx-xx-en-158
The bottom line on the onsite display displays the
pressure present, 3.9 mbar in this example.
P01-xxxxxxxx-19-xx-xx-en-159
Use "+" or "–" to switch to the "Confirm" option. The
active option is highlighted in black.
P01-xxxxxxxx-19-xx-xx-en-037
Using the "E" key, assign the value (3.9 mbar) to the
POS. ZERO ADJUST parameter. The device confirms
the adjustment and goes back to the parameter, here
POS. ZERO ADJUST (see next graphic).
P01-xxxxxxxx-19-xx-xx-en-160
Go to the next parameter with "E" .