Commissioning
31
!
Note!
• Any settings that are displayed in gray cannot be selected/cannot be changed (only notes, or
option not available/not activated).
• With the factory setting "0000" (delivery status), configuration is possible at any time. It can be
protected against unauthorized manipulation by entering a 4-digit release code. This has to be
entered when the settings are later changed, if the unit settings are to be changed by keyboard.
Tip: Make a note of your release code. Store it somewhere where unauthorized persons do not
have access to it.
• The altered settings only take effect once you return to normal operation by pressing
Z
several
times and confirming with
F
. Until this time, the unit still works with the previous data.
6.4.1 Set up - Basic settings
Settings that are not channel specific, this means date, time etc.
Procedure for basic settings:
Start
Assign unit identifier
Select the operating language from the list
Select decimal separator (comma or point)
Select temperature unit (°C, °F, K)
Select the save cycle for saving measured values
Select alarm cycle (= measured value save cycle in event of alarm limit violation)
Set the release code. Factory setting: "0000", i.e. unit setting is possible any time without code.
Make date/time settings (e.g. select region for summer time changeover).
Make settings for external memory (CompactFlash)
Set rear illumination. The "screen saver" increases the operating life of the rear illumination.
End