Fig. 21: Set up, Basic settings
"Basic settings"
menu items
Configurable parameters
(Factory settings bold marked)
Unit identifier Individual unit identifier (max. 22 characters).
!
Note!
Is also saved on the CompactFlash card.
Language Select unit operating language. Factory setting: English (GB)
Decimal separator Select in which form the decimal separator character is to be displayed. Picklist: comma, point
Temp. units Selection of temperature units. All directly connected thermocouples or resistance thermometers (RTD) are displayed in the preset
engineering units. Picklist: °C, °F, K
Save cycle Selects the cycle for normal operation (= no set point alarm) in which data is stored and displayed
Picklist: off, 1s to 1h, (1min)
!
Note!
Depending upon adjusted save cycle the available recording length changes. Tables to the typical recording length you will find in
the technical data, chapter 10.8.3.
Alarm cycle Selects the cycle for alarm condition (e.g. alarm set point violation) in which data is stored and displayed.
Picklist: off, 1s to 1h, (1min)
Release code Using this code set up access can be protected from unauthorized persons. In order to change any parameter the correct code must
be entered. Factory setting: "0", i.e. changes are possible any time.
Tip: Make a note of the code and store in a safe place.
!
Note!
This release code must be also entered in the provided PC software.
Set point code The unit is protected by a release code. If a set point code has been set up then the user can change the alarm set points after entering
this code or the unit release code (all other operating positions cannot be changed).
Factory default: "0" means that alarm set points can only be changed by entering the release code.
!
Note!
Alarm set point code and release code should not be identical!