TMT84 11 Operation using PROFIBUS
®
PA
Endress + Hauser 65
Submenu Diagnostics - Measured values - Min./max. value
This menu is only visible in the online mode.
In this menu, you can view the maximum indicators of the PV values, the two measuring inputs
and the internal reference measurement. In addition, the PV values saved can be reset.
n: Number of the Transducer Block (1-2) or the sensor input (1 or 2)
Min/max value
Menu item Parameter
"Diagnostics"
Submenu
"Measured values -
Min/max value"
Name
Parame-
ter access
Description
Primary Value n
Min.
read/
write
Min. indicator for PV is stored in the nonvolatile memory
in intervals of 10 minutes. Can be reset.
Primary Value n
Max.
read/
write
Max. indicator for PV is stored in the nonvolatile memory
in intervals of 10 minutes. Can be reset.
Sensor Value n Min. read Displays the minimum sensor value
Is stored in the nonvolatile memory in intervals of 10
minutes. Can be reset.
Sensor Value n Max. read Displays the maximum sensor value
Is stored in the nonvolatile memory in intervals of 10
minutes. Can be reset.
RJ min. value read Indicator for the minimum value to occur at the internal
reference temperature measuring point
RJ max. value read Indicator for the maximum value to occur at the internal
reference temperature measuring point