Levelflex FMP53
40 Endress+Hauser
Special installation situations
Tanks with spray ball for probe cleaning
A0014131
Mounting close to vessel wall
Mounting the probe close to the vessel wall improves the cleaning effect when a spray ball is used.
The cleaning jet is directed onto the probe via the vessel wall. As a result, the probe is also cleaned in
places the spray ball jet would normally not reach. You only require one spray ball for this probe
arrangement.
Mounting in the center of the vessel
If the probe is mounted in the center of the vessel, the use of a second spray ball may be necessary.
These balls should then be mounted to the left and right of the probe.