Levelflex FMP53
42 Endress+Hauser
Non-metal vessels
1
2
A0012527
1 Non-metal vessel
2 Metal sheet or metal flange
To ensure good measurement results when mounting on non-metal vessels, at the process
connection mount a metal sheet with a diameter of at least 200 mm (8 in) at a right angle to the
probe.