Liquiline CM442/CM444/CM448 Diagnostics and troubleshooting
Endress+Hauser 121
11.5 Adjusting diagnostic information
11.5.1 Classification of diagnostics messages
More detailed information on the current diagnostics messages displayed is provided in the
DIAG/Diagnostics list menu.
In accordance with Namur specification NE 107, the diagnostics messages are characterized by:
• Message number
• Error category (letter in front of the message number)
– F = (Failure) a malfunction has been detected
The measured value of the specific channel is no longer reliable. The cause of the
malfunction is to be found in the measuring point. Any controller connected should be set
to manual mode.
– M = (Maintenance required) Action should be taken as soon as possible
The device stills measures correctly. Immediate measures must not be taken. However,
proper maintenance efforts would prevent a possible malfunction in the future.
– C = (Function check) (No error)
Maintenance work is being performed on the device. Wait until the work has been
completed.
– S = (Out of specification) The measuring point is being operated outside its specification
Operation is still possible. However, you run the risk of increased wear, shorter operating
life or lower accuracy levels. The cause of the problem is to be found outside the measuring
point.
• Message text
If you contact the Service Department, please cite the message number only. Since you can
individually change the assignment of an error to an error category, the Service Department
cannot use this information.