Diagnostics and troubleshooting Liquiline CM442/CM444/CM448
132 Endress+Hauser
11.11 Reset measuring instrument
Sim. temperature Options
•On
•Off
Factory setting
Off
If you simulate the temperature measured value, this is
indicated on the display by a simulation icon in front of
the temperature.
Temperature -50.0 to +250.0 ˚C
(-58.0 to 482.0 ˚F)
Factory setting
20.0 ˚C (68.0 ˚F)
Set the desired simulation value.
Diagnostics/Systemtest/Reset
Function Options Info
Device reset Options
•OK
•ESC
Restart and keep all the settings
Factory default Options
•OK
•ESC
Restart with factory settings
Settings that have not been saved are lost.
Power supply Read only
• Digital Supply 1: 1.2V
• Digital Supply 2: 3.3V
• Analog Supply: 12.5V
• Sensor Supply: 24V
• Temperature
Detailed list of power supply to instrument.
The actual values can vary without a malfunction
being present.
Diagnostics/Simulation
Function Options Info