Liquiline CM442/CM444/CM448 Diagnostics and troubleshooting
Endress+Hauser 125
373 Electr. temp. M On Off High electronics temperature
--> Check ambient temperature and energy
consumption
374 Sensor check F On Off No measurement signal from sensor
-> Check the sensor connection
-> Check sensor, replace if necessary
401 Reset to default F On On Factory reset is performed
406 Param. active C Off Off --> Wait for configuration to be finished
407 Diag. active C Off Off --> Wait for maintenance to be finished
412 Writing backup F On Off --> Wait for the write process to be finished
413 Reading backup F On Off --> Wait
460 Curr. under-run S On Off Reasons
Sensor in air
Air pockets in assembly
Sensor fouled
Incorrect flow to sensor
Measures
1. Check sensor installation
2. Clean the sensor
3. Adjust assignment of current outputs
461 Current exceeded S On Off
462 Output Deviation S On Off When the current output is read back, the value
deviates from the target value displayed.
Possible reasons: Current load outside
specification, short-circuit or open current loop,
module defective
1. Check installation of current loop
2. Check module
3. Contact the Service Department
502 No text catalog F On On --> Contact the Service Department
503 Language change M On Off Language change failed
--> Contact the Service Department
530 Logbook at 80% M On Off 1. Save the logbook to the SD card and then
delete the logbook in the device
2. Set memory to ring memory
3. Deactivate logbook
531 Logbook full M On Off
532 License error M On Off --> Contact the Service Department
540 Parameter save M On Off Configuration saving has failed, --> repeat
541 Parameter load M On Off Configuration successfully loaded
542 Parameter load M On Off Configuration loading has failed, --> repeat
543 Parameter load M On Off Configuration loading aborted
544 Parameter reset M On Off Factory default successful