Technical data Liquiline CM442/CM444/CM448
166 Endress+Hauser
15.9.3 Cable specification
Cable glands mounted at the factory are tightened with 2 Nm.
15.9.4 Power consumption
CM442
Depending on supply voltage
100 to 230 V AC and 24 V AC:
Max. 55 VA
24 V DC:
Max. 22 W
CM444 and CM448
Depending on supply voltage
100 to 230 V AC:
Max. 73 VA
24 V DC:
Max. 68 W
15.9.5 Mains fuse
CM442
5x20 mm, 250 V, 4.0 A, slow-blow (T4.0A)
CM444 and CM448
Fuse is not exchangeable
Cable gland Permitted cable diameter
M16x1.5 mm 4 to 8 mm (0.16 to 0.32")
M12x1.5 mm 2 to 5 mm (0.08 to 0.20")
M20x1.5 mm 6 to 12 mm (0.24 to 0.48")
NPT3/8" 4 to 8 mm (0.16 to 0.32")
G3/8 4 to 8 mm (0.16 to 0.32")
NPT1/2" 6 to 12 mm (0.24 to 0.48")
G1/2 7 to 12 mm (0.28 to 0.48")