Operation Liquiline CM442/CM444/CM448
74 Endress+Hauser
Hold behavior Options
•Freeze
Fixed value
•None
Factory setting
Depends on the
channel:output
Freeze
The device freezes the last current value.
Fixed value
You define a fixed current value that is output at the
output.
None
A hold does not affect this current output.
Hold current
Hold behavior="Fixed value"
0.0 to 23.0 mA
Factory setting
22.0 mA
Specify which current should be output at this current
output in the event of a hold.
Measured valuedepending on theSource of data
Source of data Measured value
pH Glass Options
Raw value mV
•pH
•Temperature
pH ISFET
ORP Options
•Temperature
•ORP mV
•ORP %
Oxygen (amp.) Options
•Temperature
•Partial pressure
Concentration liquid
•Saturation
Raw value nA (only Oxygen (amp.))
Raw value μs (only Oxygen (opt. WW))
Oxygen (opt. WW)
Cond i Options
•Temperature
Conductivity
Resistance (only Cond c)
Concentration (only Cond i)
Cond c
Chlorine Options
•Temperature
•Current
•Concentration
ISE Options
•Temperature
•pH
•Ammonium
•Nitrate
Potassium
•Chloride
Path: Menu/Setup/Outputs/Current output x:y (slot:output number)
Function Options Info