Liquiline CM442/CM444/CM448 Operation
Endress+Hauser 87
10.5.3 Configuration of binary outputs
Input mode Options
•Frequency
• Parameter
Factory setting
Frequency
Frequency
Display in Hz in the measuring menu
Parameter
You subsequently determine the parameter name and
unit. These are then displayed in the measuring menu.
Parameter name
Input mode= "Parameter"
Free text Define a name for the parameter, e.g. "flow rate."
Unit of measure
Input mode= "Parameter"
Free text Determine the unit in which your parameter should be
displayed, e.g." l/min."
Lower range value
Input mode= "Parameter"
-2000.00 to 0.00
Factory setting
0.00
Your previously defined unit is additionally displayed.
The start of measuring range corresponds to a frequency
of 0 Hz.
Upper range value
Input mode= "Parameter"
0.00 to 10000.00
Factory setting
0.00
The end of measuring range corresponds to the
maximum frequency defined above. The displayed unit
is the one predefined by you.
Damping 0 to 600 s
Factory setting
0 s
The damping generates a floating average curve of the
measured value over the specified time.
1) x:y = slot no. : input number
Path: Menu/Setup/Outputs/Binary output x:y
1)
Function Options Info
Binary output Options
•Off
•On
Factory setting
On
Switching the output on/off
Signal type Options
• Static signal
•PFM
Factory setting
Static signal
Select the signal type.
Static signal
Comparable to a relay: output of a diagnostic status or a
limit contactor
PFM
You can output a measured value, e.g. the chlorine value
or the manipulated variable of a controller.
It functions as a "wear-free" switching contact that can be
used to activate a dosing pump, for example.
Path: Menu/Setup/Inputs/Binary input x:y
1)
Function Options Info