Operation Liquistation CSF33
44 Endress+Hauser
Menu/Setup/General settings
Function Options Info
Alarm delay 0 to 9999 s
Factory setting
0 s
The software displays only the errors that are
present longer than the set delay time. This
makes it possible to suppress messages that only
occur briefly and are caused by normal process-
specific fluctuations.
Device hold Selection
• Disabled
• Enabled
Factory setting
Disabled
You can enable an immediate, general hold (for
sensors) here. The function acts in the same way
as the HOLD soft key in the screens.
10.2.2 Date and time
Menu/Setup/General settings/Date/Time
Function Options Info
Set date Depends on the format Editing mode:
Day (two-digit): 01 to 31
Month (two-digit): 01 to 12
Year (four-digit): 1970 to 2106
Set time Depends on the format Editing mode:
hh (hour): 00 to 23 / 0 am to 12 pm
mm (minutes): 00 to 59
ss (seconds): 00 to 59
Extended setup
Date format Selection
• DD.MM.YYYY
• YYYY-MM-DD
• MM-DD-YYYY
Factory setting
DD.MM.YYYY
‣
Select a date format.
Time format Selection
• hh:mm am (12h)
• hh:mm (24h)
• hh:mm:ss (24h)
Factory setting
hh:mm:ss (24h)
‣
Choose between 12-hour display or 24-hour
display. Seconds can also be displayed with
the latter version.
Time zone Selection
• None
• Choice of 35 time zones
Factory setting
None
None = Greenwich Mean Time (London).
DST Selection
• Off
• Europe
• USA
• Manual
Factory setting
Off
The controller adapts the summertime/normal
time changeover automatically if you choose
European or American daylight saving time.
Manual means that you can specify the start and
end of daylight saving time yourself. Here, two
additional submenus are displayed in which you
specify the changeover date and time.